Blower for medical devices 5kPa blower
- Low noise level
- Quick acceleration response
- Compact, light weight
- Long life time
Technical Data
Motor Specification
| Unit | Value | Remarks | |
| Type of Motor | DC Brushless | ||
| Nominal Voltage | V |
12 24 |
Two types available |
| Speed Range | r/min | 0~50,000 | |
| Shaft Posture | Free | ||
| Direction of Rotation | CCW | Viewing from inlet-port side | |
| Ambient Temperature | °C | 0~50 | |
| Ambient Humidity | %RH | 15~90 | No condensation |
| Atmospheric Pressure | hPa | 770~1,013 | |
| Number of Phases | 3 | ||
| Number of Poles | 2 | ||
| Number of Slots | 6 | ||
| Type of Bearing | Ball Bearing |
Rated Values
| Unit | Value | |
| Rated Pressure | kPa | 5 |
| Rated Flow | L/min | 100 |
| Rated Speed | r/min | 42,700 |
Mechanical Data
| Unit | Value | Remarks | |
| Weight of Motor | g | 155 | Approx. |
| Acoustic Noise Level | dB(A) | 53 | 2 kPa |
| dB(A) | 62.5 | 5 kPa | |
| Rotor Ass'y Inertia | g・㎠ | 18.6 | Typical Value |
| Acceleration | rev/ms | > 110 | 0.5~5kPa: 270ms |
| Lifetime (L2) | hours | 24,000 |
5 kPa, 100L/min Calculated Value |
Electrical Data
| Unit | Value | Remarks | |
| Coil Resistance(12V) | Ω | 0.17 | 20℃ |
| Coil Resistance(24V) | Ω | 0.54 | 20℃ |
| Inductance(12V) | mH | 0.11 | 20℃ |
| Inductance(24V) | mH | 0.42 | 20℃ |